Advanced diagnostics, in accordance with NAMUR NE107.

classify alarms and warnings as ‘Maintenance Required’,

‘Check Function’, ‘Failure’ and ‘Out-of-Specification’. Cell

performance is monitored by the transmitter; indicators such as

cell impedance, rate-of-response to test gasses and changes in

calibration offset/factor are recorded and analyzed. The current

cell ‘quality’ is displayed by the transmitter as a visual indication

of the measurement confidence; providing the operator all the

information required to keep the monitor operating at peak

performance.

The Performance Log holds up to 100 time-stamped events.

When the log is full, the oldest data is overwritten by new

entries. The log contains details of measurements and

coefficients for all calibrations and accuracy checks.