Safety area classification
Class I, Division 1. Group B, C, D with Type X-Purge
Temperature Code T2 or T3
Ex Zone 1
CE 0344; II 2G; Ex db eb ib pxb IIB+H2 T2 or T3 Gb;
Conforms to EU Directives 2014/34/EU (ATEX), 2014/30/
EU (EMC) and 2014/35/EU (LVD)
EU Type Certificate LCIE 09ATEX 3006X
IECEx Certificate LCI09.0010X
Enclosure purge wait time
18.2 minutes
—
Power (hot, neutral, ground)
Voltage
100 VAC ± 10%
120 VAC ± 10%
230 VAC ± 10%
Frequency
50/60 Hz ± 10%
Power Consumption
1900 VA (maximum start-up)
1000 VA (steady-state)
—
Instrument air
Supply connection
3/8 inch tube, minimum
Supply pressure
552 to 690 kPa (80-100 psig)
Quality
Instrument grade
Clean, oil free and -34 °C, (-30 °F) dewpoint
An optional air clean up unit is available for FID burner air.
Flow rates
Startup
378 l / min (13.4 ft3/min)
Steady state
310 l / min (11 ft3 / min)
Oven Cool-down
628 l / min (22.25 ft3 / min) During vortex cool-down