Sensor power supply
Each channel is equipped with an integrated sensor power supply.
Thermocouples:
Bias voltage for isolating thermocouples
RTD: Constant current supply: 250μA …. . 1mA
Voltage: 4V
Power supply
Redundant power supply via two power inputs (decoupling diodes).
At least one power input is required for module power supply.
Supply voltage: 24 (18…. .31.2) VDC according to IEC 654-2; class DC4
Power consumption: max. 8 W
Control inputs
All channels have two logic binary inputs for the following two functions:
Reset of the maximum measured temperature
Reset latching relay
Limit blocking
The control inputs are designed for 24V logic (active low input)
Low level: 0…. .3V; high level: 13…. .48V
Input resistance: ≥30 kΩ
Current output
Nominal range:
0/4… .20 mA or 20… .4/0 mA
Open and short circuit protection
Permissible load ≤ 500 Ω
Resolution 16 bit
Optocoupler outputs
Six collector/emitter outputs
Uce = 48 VDC
Ice = 100 mA