Measurement principle
Non-dispersive infrared absorption photometers in the λ = 2.5-8 µm wavelength range can
measure up to 4 compositions using 1 or 2 beam paths and 1 or 2 receivers per beam path in one gas path or two independent gas paths.
Measure up to 4 components in one gas path or two independent gas paths with 1 or 2 receivers per beam path.
Impact Effects
Flow rate effects
Flow rates in the range of 20-100 liters/hour: within the detection range
Correlated gas effects/cross sensitivity
Analyzer configuration must be aware of sample gas composition.
Selective measures to reduce associated gas effects (optional): Installation of interference filters or filter units,
internal electronic cross-sensitivity correction of one sample component by other sample components measured by the gas analyzer.