Remarks
– The analyzer performance characteristics have been
determined according to IEC 61207-1:2010 “Expression of
performance of gas analyzers– Part 1: General”. They are
based on nitrogen as the associated gas. Compliance with
these characteristics when measuring other gas mixtures
can only be assured if their composition is known.
– All specifications refer to an optical path length (OPL) of
1 meter, tested in ABB’s test and calibration jig. However,
application-dependent variations may occur. The specific
detection limit, minimum and maximum measurement range
for a specific application will depend on the gas conditions
(pressure, temperature and gas composition) and optical
path length. Minimum measurement range, maximum
pressure and maximum temperature cannot necessarily be
realized simultaneously under all conditions.
– The maximum pressure and maximum temperature given
are physical (spectroscopic) limits.
– Applications exceeding the above given spectroscopic
limitation might be possible on request.