Instrument Description
The analyser consists of 3 separate units:
Transmitter unit with decontamination
Receiver unit with decontamination
Power supply unit
The transmitter unit consists of the laser module, the temperature stabilised diode laser,
the collimating optics and the main electronics in a coated aluminium case.
The receiver unit contains a focusing lens, photodetector and receiver electronics in a coated aluminium case.
Both the transmitter and receiver have an environmental protection rating of IP66.
The standard optical window is pressure resistant up to 5 bar (absolute pressure).
The monitor is mounted by assembling the transmitter and receiver with the supplied
purge and calibration unit and mounting to a DN50 process flange (see Figure 3-1).
Optical alignment is both simple and reliable, and the purge device prevents dust
and other contaminants from being deposited on the optical window.