Safety area classification

Class I, Division 1. Group B, C, D with Type X-Purge

Temperature Code T2 or T3

Ex Zone 1

CE 0344; II 2G; Ex db eb ib pxb IIB+H2 T2 or T3  Gb;

Conforms to EU Directives 2014/34/EU (ATEX), 2014/30/

EU (EMC) and 2014/35/EU (LVD)

EU Type Certificate LCIE 09ATEX 3006X

IECEx Certificate LCI09.0010X

Enclosure purge wait time

18.2 minutes

Power (hot, neutral, ground)

Voltage

100 VAC ± 10%

120 VAC ± 10%

230 VAC ± 10%

Frequency

50/60 Hz ± 10%

Power Consumption

1900 VA (maximum start-up)

1000 VA (steady-state)

Instrument air

Supply connection

3/8 inch tube, minimum

Supply pressure

552 to 690 kPa (80-100 psig)

Quality

Instrument grade

Clean, oil free and -34 °C, (-30 °F) dewpoint

An optional air clean up unit is available for FID burner air.

Flow rates

Startup

378 l / min (13.4 ft3/min)

Steady state

310 l / min (11 ft3 / min)

Oven Cool-down

628 l / min (22.25 ft3 / min) During vortex cool-down