Safety area classification

CSA / NRTL: Class I, Division 1; gas groups B, C, D with type Y-purge

Class I, Division 2; gas groups B, C, D

temperature code T4 – T2

ATEX / IEC / CN / KO: Zone 1: CE 0344; II2G, Ex de py IIB+H2 T4 – T2

Zone 2: CE; II3G Ex de nA nL IIB+H2 T4 – T2

Ex de px IIB+H2 T4 – T2 (optional)

With X-purge power interlock

Purge wait time: 18 minutes (Class I, Division 1 / Zone 1 area)

Power (hot, neutral, ground)

Voltage: 100 – 240 VAC

Frequency: 50-60 Hz

Power consumption: 1,200 Watts startup, 900 Watts steady-state operation

Typical, varies with installed options.

Instrument air

Supply connection: 3/8 inch tube, minimum

Supply pressure: 551.6 kPa (80 psig)

Quality: Instrument grade: Clean, oil free and -34° C, (-30° F) dewpoint

Flow rates: Steady state purge: 127-147 L/min (4.5-5.2 ft3/min) at 20° C, Y-purge types

Analytical detectors

Standard detectors: Single and multiport thermal conductivity, flame ionization, flame photometric

Third party detectors: Consult factory for availability