ABB Temperature programmed gas chromatograph PGC5000 Gas Control

Gas Control

Control Method

Closed loop PID; temperature stabilized

Number of pressure zones

1 to 5

Filtration

2 μm at inlet, provided

Inlet Pressure

Minimum

Setpoint + 69 kPa (10 psig)

Maximum

1034 kPa (150 psig)

Range

0 to 100 psig, Bubble tight, non-venting

Gauges

Electronic readout

0.01 psig resolution; Setpoint resolution:

0.01 psig

Accuracy

0 to 50 psig: 1.7%

50 to 100 psig: 2.7%

Repeatability

±0.1 psig

Allowable Gases

H2. He, N2. Air, Ar

Quality

GC Grade. an optional air clean up unit is available for FID

burner air.

Flow adjustment

Oven mounted needle valves with external adjustment

Tube fittings

316 SS Gyrolok (std.); 316 SS Swagelok (optional);

1⁄16. ⅛, ¼ inch connections

ABB PGC5000 Temperature programmed column oven

Temperature programmed column oven

Oven shell material

Stainless Steel

Internal dimensions

225.8  Diameter x 108.8 mm Depth

(8.89  Diameter x 4.285 in. Depth)

Heat

Forced air

Temperature control method

Closed loop PID

Oven temperature

+ 30 to 289 °C (Settings and display in °C only)

(86 to 35 °F)

Setpoint resolution

1 °C

Analytical columns

Metal capillary, fused Silica capillary, or packed

Stainless Steel

Column oven cooling  (When required by the application)

At the end of the analysis cycle, the controller directs

vortex-cooled air into the inner column oven to rapidly

return this oven zone to the initial temperature, allowing

the next measurement cycle to begin.

ABB PGC5000 Isothermal analytical oven

Analytical detectors

Standard detectors

Flame ionization or thermal conductivity

Both are independently heated

Isothermal analytical oven

Oven liner

Stainless Steel

Internal dimensions (W x H x D)

390 x 520 x 230 mm

(15.3 x 20.4 x 9.0 in. )

Number of valves

Standard provisions for one external liquid sample valve,

or one CP valve

Heat

Forced air

Temperature control method

Closed loop PID

Oven temperature

Ambient + 30 to 289 °C (86 to 356 °F)

(Settings and display in °C only)

Setpoint Resolution

1 °C

Temperature stability

Steady ambient

±0.1 °C (32 °F)

Ambient Range

±1.0 °C (34 °F)

ABB PGC5000 Safety area classification

Safety area classification

Class I, Division 1. Group B, C, D with Type X-Purge

Temperature Code T2 or T3

Ex Zone 1

CE 0344; II 2G; Ex db eb ib pxb IIB+H2 T2 or T3  Gb;

Conforms to EU Directives 2014/34/EU (ATEX), 2014/30/

EU (EMC) and 2014/35/EU (LVD)

EU Type Certificate LCIE 09ATEX 3006X

IECEx Certificate LCI09.0010X

Enclosure purge wait time

18.2 minutes

Power (hot, neutral, ground)

Voltage

100 VAC ± 10%

120 VAC ± 10%

230 VAC ± 10%

Frequency

50/60 Hz ± 10%

Power Consumption

1900 VA (maximum start-up)

1000 VA (steady-state)

Instrument air

Supply connection

3/8 inch tube, minimum

Supply pressure

552 to 690 kPa (80-100 psig)

Quality

Instrument grade

Clean, oil free and -34 °C, (-30 °F) dewpoint

An optional air clean up unit is available for FID burner air.

Flow rates

Startup

378 l / min (13.4 ft3/min)

Steady state

310 l / min (11 ft3 / min)

Oven Cool-down

628 l / min (22.25 ft3 / min) During vortex cool-down

ABB PGC5000 TP Gas Chromatograph

Description

An isothermal oven contains the temperature programmed

oven assembly with a detector, either a Flame Ionization

Detector (FID) or Thermal Conductivity Detector (TCD). The

detector is mounted on top of the temperature programmed

oven, and it is integrated with the TP oven in order to provide

an optimal analytical flow path and eliminate potential cold

spots. The sample valves, either a Liquid Sample Valve (LSV)

or a Continuous Performance (CP) valve, can also be installed

in the isothermal oven compartment. An optional air

cleanup / methanizer unit may be used. Special columns

(capillary or packed type) are installed inside the temperature

programmed oven. The columns contain special packing

material which separates the compounds to be analyzed so

that they enter the detector in a predictable sequence.

Specialty analysis for separation of wide boiling range

compounds. The PGC5000 Temperature Programmed (TP)

Gas Chromatograph uses component separation to analyze

gas or liquid samples. The analyzer operates unattended,

automatically sampling and analyzing process streams.

ABB PGC5000 Temperature Programmed (TP) Gas Chromatograph

Introduction

Specialty analysis for separation of wide boiling range

compounds. The PGC5000 Temperature Programmed (TP)

Gas Chromatograph uses component separation to analyze

gas or liquid samples. The analyzer operates unattended,

automatically sampling and analyzing process streams.

Physical

Environmental (enclosure)

Protected from weather:  IP 54. (NEMA Type 13) equivalent

Ambient temperature range

0 to +50 °C (32 to 122 °F)

Humidity

95% relative humidity, non-condensing

Dimensions (W x D x H)

496 x 340 x 1175 mm

(19.5 x 13.4 x 46.3 in.)

Weight

73 kg (160 lb) (minimum)

Mounting

Wall: 32 mm (1.3 in.) from wall with brackets

Floor: Optional wheeled dolly

EMI/RFI considerations

Conforms to Class A industrial environment

Electrical entries

Left side

Pneumatic entries

Right side

Sample entries

Liquid

Right side, 1 each model 791 LSV

Gas

Bottom and/or right side

Vents

Bottom and/or right side

ABB PGC5000A gas chromatographs Safety area classification

Safety area classification

CSA / NRTL

Class I, Division 1; gas groups B, C, D with type Y-purge

Class I, Division 2; gas groups B, C, D Temperature Code T4

ATEX / IECEx/ CN / KO

Zone 1: CE; II2G, Ex ib py IIB+H2 T4 Gb

Zone 2: CE; II3G Ex nA IIB+H2 T4 Gb

CU TR: INMETRO

Optional X-purge power interlock Ex d ib px IIB+H2 T4

Purge wait time

18.2 minutes (Class I, Division 1 / Zone 1 area)

(hot, neutral, ground)

Power (hot, neutral, ground)

Voltage: 100 to 240 VAC

Frequency: 50 to 60 Hz

Power consumption: 120 VA startup and steady-state

operation (variable with installed options)

Instrument air

Supply connection: ¼ inch tube, minimum

Supply pressure: 414 kPa (60 psig) minimum

Quality: Clean, oil free and –34° C, ( –30° F) dew point

Instrument grade:

Flow rates – steady state purge

2 to 25 l / min (0.75 to 0.78 ft3 / min) at 20° C,

Y-purge types

ABB PGC5000 Series gas chromatographs PGC5000A

Application flexibility

• PGC5000B Smart Ovens target basic applications

with a fixed set of features

• PGC5000C Smart Ovens target complex

applications requiring multiple detectors for high

application density

• PGC5007 oven targets flare gas monitoring and

total sulfur process applications

• PGC5009 oven designed for fast temperature

programmed simulated distillation analysis

Specification

Master controller

Environmental (enclosure)

Protected from weather

IP 54. (NEMA 3 equivalent)

Ambient temperature range

0 to 50° C (32 to 122° F)

Humidity

95% relative humidity, non-condensing

Dimensions (W x D x H)

596.9 mm x 419.1 mm x 256.5 mm

(23.5 in. x 16.5 in. x 10.1 in.)

Weight

20 kg (44 lbs) (minimum, configuration dependent)

Mounting

Wall: 33 mm (1.3 in.) from wall with brackets

Floor

Optional dolly with casters

EMI / RFI considerations

Conform to class A industrial environment

Electrical entries

Top

ABB PGC5000 Series Multiple configurations

Features

Ease of use

• Graphics based touch screen HMI

Multiple configurations

• Supports multiple Smart OvenTM configurations,

•  Up to 4 PGC5000B ovens

• Up to 2 PGC5000C ovens

• One PGC5000C with up to 2 PGC5000B ovens

Communication interfaces

• Listing_01: Nempello qui tem dolestiuria

– Listing_02:, qui ipsa volupta

• Ethernet, OPC, MODBUS, 4 to 20mA analog

outputs, VistaNET 2.0 compatible

– Nempello qui tem dolestiuria dolenimus

– Meniasperrum quuntem

Enhanced Data Storage

– Cia iumendaest, qui ipsa volupta

• Nempello qui tem dolestiuria dolenimus

• Up to 7 days of chromatograms and reports via

removable SSD card

• Meniasperrum quuntem

Standard inputs / outputs

• Eight isolated 4 to 20mA outputs – Four digital outputs

– Dedicated purge alarm

Dolupta arum velia volenimus disci temqui offic

tentiorrum quis atquiberis doluptatem non pore,

Sendus dolut que illorio quod quundem

quunduciatia pre modis quam quunduciatia premodis

– Dedicated common malfunction alarm

• Two additional configurable outputs

ABB PGC5000 Series gas chromatographs PGC5000A

The new standard in industrial gas chromatographs

Measurement made easy

Overview

The PGC5000A Generation 2 master controller

provides all analyzer system control functions and

coordinates internal and external data activities for

all of the PGC5000 Series ovens. The new controller

also supports multiple smart oven configurations to

maximize application flexibility while minimizing

space and utility requirements. The Generation 2

controller is designed with a Real Time (embedded)

Operating System (RTOS) to guarantee critical

system uptime, security and deterministic transfer

of data. The Human Machine Interface (HMI)

includes a 10.4 inch true color touchscreen SVGA

display, universal keypad and touch-pad mouse.