SPECIFICATIONS

Supported sensors

Currently available :Compatible with a range of sensors and signal conditioners using

2-wire current transmission and 3-wire voltage transmission

Dynamic input channels

Number of independent channels : 2

Measurement range

•Voltage (AC+DC) : ± 3 or ± 30 VPEAK

•Current : 15 or 150 mA

Frequency bandwidth (−0.1 dB) : 0.1 Hz to 19 kHz

•Analog high-pass filter :An optional (software configurable) analog high-pass filter can be

added to the AC path to configure the high-pass cutoff frequency

(−3 dB) as 0.1. 1.0 or 3.0 Hz.

This filter can also be disabled to allow DC-coupling of the input.

Input impedance

•Voltage : ≥ 100 kΩ, between the differential (high and low) inputs

•Current : 202 Ω ± 3 Ω, between the differential (high and low) inputs