ABB EasyLine Continuous Gas Analyzer EL3020 Temperature effects
Temperature effects
Ambient temperature within the permissible range
– At zero point: ≤ 2 % of range per 10 °C – At sensitivity without thermostat:
≤ 3 % of the measured value per 10 °C – Sensitivity with thermostat (optional):
Sensitivity with thermostat (optional): ≤ 2 % of measured value per 10 °C
Thermostat temperature = 55 °C
Barometric influences – At zero point: No influence – Sensitivity during pressure correction by means of integral pressure sensor
Per 1 % barometric pressure change: ≤ 0.2 % of measured value
Barometric pressure variation
If a hose is used for the internal gas line, the pressure sensor is located in the sampling gas line.
If a hose is used for the internal gas line, the pressure sensor is routed externally through the hose.
Operating range of the pressure sensor: pabs = 600-1250 hPa
Power supply influences
Voltage and frequency within the permissible range: no effect