Emerson 1066 Liquid Analysis Fieldbus pH/ORP Transmitter
The terminals are listed below:
TB3 RTD INPUT TERMINALS: The leads of the 3-wire RTD should be wired as shown.
If a 2-wire RTD is used, the RTD return and RTD sense terminals must be jumpered to avoid an open RTD sense wire warning.
TB2 Reference Electrode and Solution Ground: The reference electrode lead and its shield
and the solution ground lead should be grounded as shown. If the sensor does not have a solution ground, there are two options:
1. The reference voltage input and solution ground can be jumpered.
If this is done, the reference impedance will read a constant value of 0 kilohms.
2. The second method is to turn on the solution ground terminal and set the Reference
Impedance parameter (Reference Z) in the program menu (see Section 7.3.7) to High,
thus turning off the reference impedance measurement.
If the solution ground terminal is left open without doing so, a high reference impedance fault alarm will continue to sound.
TB4 Preamplifier Power: The power wire from the pH sensor or preamplifier in the junction box is connected to this terminal to supply power to the preamplifier.
TB1 pH Electrode Input: The pH electrode lead and its shield are located on this terminal as shown.
Smart pH Sensor: The smart pH sensor has a ground wire (not to be confused with the solution ground wire)
that should be connected to the enclosure ground as shown in the power wiring diagram