Specifications

Inputs per module: 8

Thermocouple types: J, K, T, E, R, S, N, and mV

Isolation: 1500 VDC channel-to-backplane

Input overrange protection:

30 VDC or VAC continuous

120 VAC for 10 Sec

Measurement ranges:

J: -210oC to 760oC (-350oF to 1400oF)

K: -270oC to 1372oC (-450oF to 2500oF)

T: -270oC to 400oC (-454oF to 752oF)

E: -270oC to 1000oC (-454oF to 1832oF)

R: 0oC to 1768oC (32oF to 3214oF)

S: 0oC to 1768oC (32oF to 3214oF)

N: -260oC to 1300oC (-436oF to 2372oF)

mV: -55 to +55mV

Millivolt input impedance:

4.4MΩ DC, >10KΩ @60 Hz

Open Thermocouple Detect: less than 1 Sec

Data presentation:

With scaling enabled = 0 – 32000 (temp.)

-32000 to +32000 (mV)

With scaling disabled = oC or oF x10 (temp.)

-10000 to +10000 (mV)

Measurement units: Degrees C or F

selectable by module

Digital Filtering Time Constant: 80 mSec

Update time:

9 mSec all channels

Repeatability: +/-1oC, +/-2oF for all

thermocouple types

Accuracy: Types

J: +/-1oC from 0oC to 60oC

K: +/-1oC from 0oC to 60oC

T: +/-1oC from 0oC to 60oC

E: +/-1oC from 0oC to 60oC

R: +/-2oC from 0oC to 60oC*

S: +/-2oC from 0oC to 60oC*

N: +/-1oC from 0oC to 60oC

mV: +/-0.5% full scale or +/- 500μV

ADC Resolution: 16 bits

Common Mode Rejection: >110 db @ 60 Hz

Normal Mode Rejection:

>60 db @ 60 Hz, >50 db @ 50 Hz

Connector: Removable

Wire Gauge: 14-22 AWG

Backplane Power: 1.0 Watt (maximum)

Module Size: Single-wide

Operating Temperature: 0o to 60oC (32o to 140oF)

Storage Temperature: -40o to 85oC (-40o to 185oF)

Relative Humidity: 5% to 95% (non-condensing)

Agency Approvals Pending:

UL, ULC, Class 1-Div. 2. CE

Shipping Weight: 1.5 lb. (0.68 Kg)