Measuring principle

The OXITEC® 5000 O2 analyzer system consists of an InSitu probe

which is installed in a duct to measure non combustible process

gases and of an electronic unit for voltage and gas supply, as well

as for signal processing.

The oxygen sensor is at the tip of the probe and is regulated to

800 °C and works on the zirconium oxide principle of measurement.

Here, a mV signal between the reference gas side of the sensor

(inside, instrument air 20.95% O2) and the measured gas side is

measured, which depends logarithmically on the ratio of oxygen

partial pressures on both sides of the sensor.

The mV signal is converted according to the Nernst equation into

oxygen partial pressure within the process gas, whereby the O2

concentration is determined in the process gas. Gas-tight separation

of reference air and process gas is of particular importance.

Should combustible components such as CO or H2 be present in the

sample gas, they will react with oxygen at the sensor surface and can

reduce the measured value.