ENOTEC OXITEC® 5000+ Disruption of the process

Disruption of the process

The analyzer system has to be kept in operation also in the event of

the process being disrupted or if the plant is powered off temporarily

(e.g. at night or during the weekend).

Frequently cooling down and heating up of the probe results in thermal

stress of the hot probe parts (heater, thermocouple and sensor) and

reduces their product life. ENOTEC will not accept any responsibility for

resultant damage.

Storage instructions

ENOTEC equipment and spares are to be stored in a dry and ventilated

environment at temperatures between -40 °C to +80 °C (-40 °F to 176 °F).

Paint fumes, silicone sprays, etc. must be avoided in the storage environment.

Name plates

The name plate contains information about the line voltage, the nominated

current, frequency, protection class, year of manufacture, serial number, order

number and system order code.

The system order code refers to information which is detailed in the system

test report and supplied with the system.

ENOTEC OXITEC® 5000 O2 analyzer system

Measuring principle

The OXITEC® 5000 O2 analyzer system consists of an InSitu probe

which is installed in a duct to measure non combustible process

gases and of an electronic unit for voltage and gas supply, as well

as for signal processing.

The oxygen sensor is at the tip of the probe and is regulated to

800 °C and works on the zirconium oxide principle of measurement.

Here, a mV signal between the reference gas side of the sensor

(inside, instrument air 20.95% O2) and the measured gas side is

measured, which depends logarithmically on the ratio of oxygen

partial pressures on both sides of the sensor.

The mV signal is converted according to the Nernst equation into

oxygen partial pressure within the process gas, whereby the O2

concentration is determined in the process gas. Gas-tight separation

of reference air and process gas is of particular importance.

Should combustible components such as CO or H2 be present in the

sample gas, they will react with oxygen at the sensor surface and can

reduce the measured value.

ENOTEC COMTEC® 6000 GasEx analyzer Technical data

Technical data

Process gas temperature Max. 500 °C – max. 1400 °C with cooling

protection tube

Measuring principle O₂ Zirconium oxide|MXP

Immersion depth/material 403 mm to 863 mm (others on request) /

Stainless steel (SS316TI)

Measuring principle COe MXP (Mixed Potential)

Ambient temperature Probe: -40°C to +70°C | Electronics: -20°C to

+55°C

O₂ – Measuring ranges Two freely selectable measuring ranges: 0-2 vol.%

to 0-21 vol.% (others on request)

COe – Measuring ranges 0 to 5000 ppm

Housing Probe: IP66 | Electronics: IP66

Type of protection II 2G Ex db IIC T3 Gb

ENOTEC AQUATEC® 1000 Technical data

Technical data

Process gas temperature Max. 200 °C (with activated carbon filter)

Max. 400 °C (with ceramic filter)

Measuring principle O₂ Zircon oxide

Immersion depth/material Max. 615 mm (others on request) / Stainless steel

(SS316TI)

Ambient temperature Probe: -40°C to +80°C | Electronics: -20°C to +55°C

O₂ – Measuring ranges Two freely selectable measuring ranges: 0-2 vol.% to

0-25 vol.% (others on request)

H₂O – Measuring ranges 0 to 100%

Housing Probe: IP65 | Electronics: IP66

ENOTEC CEMTEC® hot/wet Only CEMTEC guarantees 95 % availability of kiln gas values

Only CEMTEC guarantees 95 % availability of kiln gas values.

CEMTEC requires only one maintenance cycle per year, due to the highly

effective self-cleaning and monitoring features of this robust system.

All drives in the CEMTEC system are pneumatically operated, which is a

distinct advantage compared to electrical drives.

• The pneumatic motor for insertion and retraction, which automatically

removes the probe with a force of 2 tons.

• The pneumatic turning actuator with 700 Nm prevents the probe from

sticking in the process and material build up on the probe.

• The pneumatic plunger motor regularly keeps the gas entrance free

from clogging.

• A 1000 l compressed buffer tank is a permanent energy reserve for the

probe system. Should any event potentially threaten the probe, e.g.,

coolant under/over-temperature, under/over-pressure or loss of power

(an electrical blackout in the plant), the probe will automatically retract

from the kiln.

In addition, the probe tube is purged periodically by the blaster tank,

removing all residual dust from the internal large probe filter tube.

ENOTEC AQUATEC® 1000 Developed to optimize the drying process

Developed to optimize the drying process

Measuring range up to 100% H₂O and O₂

Robust probe design for high drying temperatures

Dual system configuration allows measurement in special drying processes

Drying processes are among the most energy-intensive steps in manufacturing

processes, whether it’s the production of plasterboard or the processing of

tobacco plants. Fast and reliable measurements during the drying process

reduce the time required and allow for optimized energy use.

Due to its in-situ measurement method, the AQUATEC 1000 analyzer is superior

to conventional systems using extractive methods in terms of measurement

speed.

Sensors from ENOTEC have already proven themselves in countless applications

worldwide.

ENOTEC ENSITU 7000 enables fast and continuous measurement

ENSITU 7000 enables fast and continuous measurement and

is capable of withstanding flue gas temperatures up to 600°C and high

dust loads. Versatile and robust, this new ENOTEC analyzer is the ideal

solution for process control in small to medium-sized boilers using all

applicable fuels, such as gas, oil, and coal. The ENSITU 7000 requires

no calibration or instrument air and is ready for immediate use.

The ENSITU 7000 measures in situ: The flue gas flows directly through

the large filter with a V-shield, so the O₂ sensor reacts immediately to

changes in oxygen concentration, eliminating the need for flue gas

extraction or downstream gas treatment.

Features: Plug & Play Operation

• Innovative operating concept

• Ready for immediate use

• Easy integration into existing boiler concepts

• Electronics detachable from the probe via a robust industrial connector

ENOTEC COMTEC® 6000 StaubEx Technical data

Technical data

Process gas temperature max. 600 °C

Measuring principle O₂ Zirconium oxide|MXP

Immersion depth/material 540 mm to 960 mm (others on request) /

Stainless steel (SS316TI)

Measuring principle COe MXP (mixed potential)

Ambient temperature Probe: -20°C to +70°C | Electronics: -20°C to

+55°C

O₂ – Measuring ranges Two freely selectable measuring ranges: 0-2 vol.%

to 0-21 vol.% (others on request)

COe – Measuring ranges 0 to 5000 ppm

Housing Probe: IP66 | Electronics: IP66

Type of protection II 2D Ex tb IIIC T133°C/141°C Db

ENOTEC ENSITU® 7000 Developed for maximum cost efficiency

Developed for maximum cost efficiency

• Plug & play

• Electronics installed directly on the sensor

• Instrument air is NOT required

• All functions controllable and storable via ENOTEC Remote APP

Technical data

Process gas temperature max. 600 °C

Measuring principle O₂ Zircon oxide

Immersion depth/material 313 mm to 993 mm (others on request)

/ Stainless steel (SS316TI)

Ambient temperature Probe: -40°C to +55°C | Electronics: -40°C to +70°C

O₂ – Measuring ranges Two freely selectable measuring ranges: 0-2 vol.%

to 0-25 vol.% (others on request)

Housing Probe: IP65 | Electronics: IP66

ENSITU 7000. ENOTEC’s affordable entry-level product, is based on

decades of experience. Development, design, and manufacturing take

place exclusively at the German headquarters – from the sensor to the

software.

ENOTEC CEMTEC® Reliable and maintenance-free gas extraction system for extreme combustion processes

Cross-section: Gas inlet CEMTEC probe tube

The CEMTEC is self-cleaning, which enables continuous measurement and

thus 95% availability during the process. In addition to the in-situ sensors

for oxygen and COe analysis, the extractive analyzers are supplied with

dust-free sample gas for the determination of, for example, SO₂, NOₓ, or

CO₂. Thanks to mechanical cleaning, the uninterrupted extraction and

analysis of process gas is possible over extended periods: The filter/seal at

the gas inlet continuously breaks down deposits and blockages, ensuring

an uninterrupted gas supply. The patented swivel drive rotates the probe

at programmable intervals, thus preventing material from accumulating on

the probe tube. Together with the powerful pneumatic drive motor, the

probe can perform a controlled retraction at any time.

Technical data

Process gas temperature ≤ 1400 °C

Immersion depth/material up to 4.5 m

Dust loading up to 2000 g/m³

Probe cooling method Air/coolant or water/coolant

Probe operation 100% pneumatic drives (extension and retraction, rotation,

plunger and flushing)